Hematopoietic Stem/Progenitor Cell Enumeration in Apheresis Collections Using a New Microfluidic Cartridge-Based Flow Cytometry Platform

Automating flow cytometric analysis enables all levels of technicians to generate accurate and precise cell enumeration and phenotyping results in real time, providing actionable in-process information.

Hematopoietic Stem/Progenitor Cell Enumeration in Apheresis Collections

Filename
Accellix-2025-ISCT-HSC-Enumeration-Mount-Sinai-Poster.pdf
Size
1 MB
Format
application/pdf

Summary

  • A new method for CD34+ cells phenotyping and enumeration in fresh HPC, A samples is in advanced stages of evaluation using the Accellix Automated Microfluidic-Based Flow Cytometry platform.
  • New gating strategy, methods of analysis and reportables have been established in line with current clinical standards.
  • Method comparison between the Accellix and predicate flow cytometry platforms demonstrate:
  • Comparable measurements of cell viability for both CD45+ and CD34+ cells.
  • Nearly identical frequencies of viable CD34+ cells showing strong correlations and concordance.
  • No significant difference and strong correlations between the absolute numbers of viable CD34+ cells.
  • Acceptable linearity in HPC, A serial dilutions and CD34+ cells spiking assays; reproducibility and inter-operator variation had a coefficient of variation of ≤10% [(n=10), data not shown].
  • Manual and automated analyses of the results obtained on the Accellix platform are comparable thus warranting transition to method validation and implementation.
  • Preliminary results indicate feasibility and accuracy of Accellix method for dual CD34+/CD3+ phenotyping and enumeration in a single assay (data not shown).