Skip to main content

Automating Flow Cytometry for Objective & Reproducible Cell Therapy Quality Control

   

This poster presents the ACCELLIX™ solution and how it demonstrated performance across 20+ Leukapheresis samples & control materials.

Summary

  • The ACCELLIX™ Cell Phenotyping Platform automates sample preparation, voltage calibration, spectral unmixing, data acquisition, and data analysis.
  • T Cell and TBNK assay cell subsets were clearly and consistently identified by assay-specific autoclassification software across 30 unique leukapheresis samples, 3 lots of CD Chex Plus®, 15 ACCELLIX™ instruments, and cell concentration levels ranging from 1 to 50 million cells per mL.
  • Coefficients of variance (CVs) for 5 replicates (5 instruments) across all samples ranged from 0.81% to 7.69%.
  • Comparing manual and automated analysis results for each cell subset, the coefficients of determination (R2 values) are all at or above 0.9 and slopes range from 0.98 to 1.03, demonstrating excellent concordance and no systematic bias.

The ACCELLIX™ platform has the potential to improve cell therapy QC efficiency by providing a simple and rapid workflow that minimizes process variability and subjective data analysis.

ACCELLIX™ Reproducible QC Poster

Download the Poster

Complete the form below to access the poster, which will be sent directly to your email after submission. By providing your information, you agree to receive additional emails from bioMérieux about cell therapy quality control. 

To find out more about the management of your personal data and to exercise your rights, see our privacy policy.