In this study, the Most Probable Number (MPN) was used to demonstrate that the LoD of the SCANRDI® CELL-BURST is not significantly different from the LoD of a traditional plate counting method on 10 compendial strains. To show the equivalency, the proportion of positive results was compared between both methods for all strains and all dilutions with a non-inferiority test of Farrington-Manning, with a 20% margin.